1. IMPORTANT! This software tested with: TELEMED ultrasound systems ---------------------------- - ArtUs scanners - ArtUs Compact scanners - MicrUs scanners - MicrUs Pro scanners - SmartUs scanners - ClarUs scanners Can be used with: Obsolete ultrasound systems ---------------------------- - LS128 scanners - LS64 scanners - Echo Blaster 128 scanners - Echo Blaster 64 scanners 2. RECOMMENDED COMPUTER REQUIREMENTS ---------------------------- Hardware ---------------------------- - Microsoft Windows compatible desktop/notebook/Tablet PC - Intel chipset based motherboard with integrated USB controller: USB 2.0 or USB 3.0: MicrUs scanner MicrUs Pro scanner SmartUs scanner ClarUs scanner LS128 scanner LS64 scanner Echo Blaster 128 scanner Echo Blaster 64 scanner USB 3.0: ArtUs scanner - CPU i5/i7/i9 1.8 GHz or faster - 2 GB of RAM or more - 100 MB free hard disk space - Monitor with 1024x768 resolution support or higher, anti-reflective surface, IPS ---------------------------- Operating system ---------------------------- - Windows 7, Windows 8, Windows 10, Windows 11 with latest updates (all versions 32/64-bit) 3. HISTORY Ver 2.2.1 2024 September 1 + For ArtUs scanners fixed AGC Power compensation. Ver 2.2.0 2024 April 10 + Changed max PW SVL setting for ArtUs probe C15-7N30-A4. + Added support of probes C15-7N30-A4, P6-3S7-A, MC10-4N12-A4, L13-4H55-A15, L15-6N25-A4, BIPC8-4N10-A15 and BIPL10-4N60-A15 for Artus scanners. + Improved image resolution and contrast (iContrast). Ver 2.1.3 2022 November 1 + Synchronization with external devices is turned off during probe elements test for ArtUs and SmartUs scanners. + Improved probe check for SmartUs scanners. + Added USB specification to log output. + "Deep freeze" is implemented for MicrUs Pro scanner series. + Added bioeffect indices tables of probes BIPC9-4H10-A15, BIPL12-5H70-A15 for Artus scanners. + Added support of ArtUs2H scanners. Ver 2.1.2 2021 August 31 + (00299) Added support for MicrUs Pro Rev.B. + (00300) Added support of probes BIPC9-4H10-A15, BIPL12-5H70-A15 for ArtUs devices. Ver 2.1.1 2021 May 04 + (00276) Fixed bug with B-mode data transfer in B+M mode scanning. Ver 2.1.0 2021 April 16 + Improved RF ultrasound data processing. + Improved accuracy of Artus scanner programming. + Added PW mode settings for all probes supported by MicrUs rev.E scanners. Ver 2.0.2 2020 November 17 + Added workaround code for ArtUs scanners to fix problem with Super Speed USB data transfer on Windows 7 OS. + Micrus rev.E new device ID. + Improve focusing in Artus device + Changed apodization setting file for ArtUs devices. + Added bioeffect indices table for LF11-5H60-A3 for ArtUs scanner + Added support of probe MC11-6R6L-9 for SmartUs scanners + Added support of probes L8-4L40S-14, C7-3R60S-14, MC7-3R15S-14 for MicrUs scanners Ver 2.0.1 2020 September 25 + Added probe elements check for SmartUs rev.E devices. + Updated settings of probes LF11-5H60-A3, L12-5N40-A4 for ArtUs scanners. + Added support of SmartUs rev.E devices. + Added bioeffect indices table for probe LF9-5N60-A3 for ArtUs scanners. + Improved beam time limitation algorithm for MicrUs Pro devices (overheating control). + Added support for MicrUs Pro-L40S, -L40N, -C60S devices. + Added support of probe BIPCD10-5R10L-5 for SmartUs rev.C devices. + Added bioeffect indices table of probe L12-5N40-A4 for ArtUs scanners. + Added support of probe BIPL10-4H70-A4 for ArtUs scanners. + Implemented probe test surface procedure for ArtUs scanners. + Implemented probe test surface procedure for SmartUs scanners. + Implemented Power Doppler modes for ArtUs scanners. + Added management of compound scanning type for RF scanning for ArtUs scanners. Ver 2.0.0 2019 August 21 + Added new synchronization type, ArtUs scanner. + Improved probe recognition, LS64 scanner. + Improved fan/cooling control, ArtUs scanner. + Added bioeffect indices for probes, ArtUs and SmartUs scanners. + Added possibility to load custom TFC filter, ArtUs scanner. + Increased maximum scanning depth, ArtUs scanner. + Added 320Hz and 400 Hz PRF values, color doppler modes, ArtUs scanner. + Added multibeam for Color Doppler modes, ArtUs scanner. + Added support of various probes, ArtUs scanner. + Added support of PW mode, ArtUs scanner. + Updated indication of deepest focus position, MicrUs scanner. + Changed power management when powered from external PS, MicrUs scanner + Changed installation scenario, all scanners. + Changed Dynamic Range management, ArtUs scanner. Ver 1.17.3 2018 September 12 + Fixed bug in MicrUs Duo installation script. + Added support of Artus devices based on BULK USB data transfer protocol. + Added support of NGS featured ArtUs devices. + Added support of probe codes 0F for MicrUs rev.C devices. + Added support of test probe for ArtUs devices. + Changed FPGA data file for ArtUs device. + Impruved scanning mode change procedure for ArtUs devices. + Added support of ArtUs with firmware rev.00.10. + Added support of probes L15-7H40-A5, L18-8H30-A5 for ArtUs scanners. + Added support of probe codes 0E, 0F for MicrUs rev.D devices. + Added support of probe code 67 for SmartUs rev.C devices. + Fixed bug with MicrUs Duo installation. + Added 128 elements test probe definition for MicrUs rev.D scanner installation script. + Changed management of B-mode scanning type (avoid to change scanning type if scanning mode is not B). + Fixed bug in ArtUs scanner with doppler window left shift after B-mode format change. + Fixed bug in ArtUs scanner with wrong frame buffer size declaration. + Changed internal device temperature management for SmartUs devices. + Updated settings of all probes for ArtUs devices. + Fixed bug with drivers mismatch for SmartUs rev.B scanner. + Added support of probes P4-2S20-A3, C5-2N60-A3, L12-5N40-A3 for ArtUs devices. + Changed power settings for all probes of ArtUs devices. + Added support of probe L14-3iN for ArtUs devices. + Updated settings of probes for ArtUs devices. + ArtUs driver code changed to allow to apply custom AGC and TGC power compensation values for each B-mode frequency. + Fixed bugs in multibeam mode scanning for ArtUs devices. + Changed supply power management of AtrUs devices. + Changed transmit power calculation for ArtUs devices. + Updated MicrUs devices driver to output synch signals by default. + Added support of probe L18-8L40H-5 for SmartUs rev.C devices. + Fixed bugs in RF data transfer for ArtUs devices. + Added support of probe L12-5N60-A2 for ArtUs devices. + Changed processing of following controls of ArtUs devices: Dynamic Range, TGC, AGC, ADC-AFE. + Improved FPGA doppler functionality for ArtUs devices. + Added processing of new firmware variables for ArtUs devices. + Added support of 4M SmartUs devices. + Changed settings of probe L12-5L40N-2 for ClarUs devices. + Added bioeffect indices table of probe L12-5L40N-2 for ClarUs devices. + Fixed bug with several applications usage of ArtUs devices simultaneously (previous beta). + Added sync output signals management for ArtUs devices. + Added support of probe LV8-5N60-A2 for ArtUs devices. + Drivers are signed with Microsoft certificate. + Changed settings of probe LW12-5L50N-3 for MicrUs rev.C, D devices. + Changed settings of probe C6-2R50HI-5 for SmartUs rev.C devices. + Fixed bug with focus programming in wide view mode for ArtUs devices. + For ArtUs devices multibeam mode is disabled for Color Doppler scanning modes. + Fixed bug with phased probe focus programming for ArtUs devices. + Changed settings of probe C6-2R50HI-5 for SmartUs rev.C devices. + Added support of probe C6-2R50HI-5 for SmartUs rev.C devices. + Changed settings of probes L15-6N25-M3, L12-5N40-M3, L12-5N60-M3, L12-5N50-M3 for MicrUs rev.D devices. + Changed range of line density property for MicrUs rev.C, D devices. + Added support of probes L15-6N25-M3, L12-5N40-M3, L12-5N60-M3, L12-5N50-M3 for MicrUs rev.D devices. Ver 1.17.2 2017 October 20 + Fixed bug in installation script of MicrUs rev.C device driver. + Added possibility to scan in ITHI mode for 3D mode for K series devices. + Added support of probe L13-4L55H-4 for SamrtUs rev.C devices. + Fixed MicrUs rev.C,D 64-bit system issue. + Changed MicrUs rev.C, MicrUs Duo, MicrUs rev.D installation management. + Added support of probe LAP12-5L40N-4 for SmartUs rev.C devices. + Added support of probe L15-6N25-M4 for MicrUs rev.D devices. - Added support of probe L12-5N40-M4 for MicrusDuo rev.C,D devices. - Probes for ArtUs devices are renamed accodring new naming specification. - Changed settings of probes MC10-4R12N-4 and L12-5L60N-2 for SmartUs rev.C,D devices. - Added support of probe P7-3L5L-9 for SmartUs rev.C, D devices. - Added support of ArtUs research revision device support. - Added support of probe C6.5/15/64-3 for LS64 devices. - Changed settings of probe C3.5/20/64-3 for LS64 devices. - Added support of probe BIPC9-4R9L-7 for SmartUs rev.C,D devices. - Changed settings of probe MC9-4R15S-3 for MicrUs rev.C,D devices. Ver 1.17.1 2017 May 15 - Changed settings of probe HL9.0/60/128-2 for LS128 devices. - Changed settings of probe L12-5L60N-2 for ClarUs devices. - Changed settings of probe C5-2R60HI-5H for ArtUs devices. - Changed settings of probe LV8-5L60N-2 for SmartUs devices. - Added support of probe MCC4-2R20S-3 for MicrUs devices. - Added support of probe L14-5L40N-4 for ClarUs devices. - Added support of probe EC6.5/10/128-3 for Echo Blaster 128 devices. - Fixed bugs in M-modes scanning for ArtUs devices. - Changed settings of probe L18-10L30H-4H for ArtUs devices. - Added support of compound mode for convex probes for ArtUs devices. - Changed apodization filters for ArtUs, SmartUs rev.B, rev.C, rev.D devices. - Added support of probe MCC10-5R10S-3 for MicrUs Duo rev.C,D devices. - Added support of probe MCC10-5R10S-3 for MicrUs rev.C devices. - Fixed bug with wrong power supply management for SmarUs devices. - Fixed bugs in wideview mode scanning for Artus devices. - Added support of prbe ID 0086 for SmartUs rev.B devices. - Changed bioeffect indices table of probe CLA6.0/10/148 for SmartUs rev.B,C device. - Updated FPGA for ArtUs devices. - Added support of phased array probes for Artus devices. - Updated settings of Artus device probes. - Added support of probe LD12-5L40S-3 for Micrus Duo devices. - Added support of back curved convex probes for SmartUs devices. - Added support of probe CA11-4R85H-4 for SmartUs rev.C,D devices. Ver 1.17.0 2016 December 19 - Updated settings of probe CLA6.0/10/148 for SmartUs rev.B,C devices. - Updated bioeffect indices of probes L18-10L30H-4 and P5-1L15SI-6 for SmartUs rev.C device series. - Scaning parameters of probe L18-10L30H-4 for SmartUs rev.C device series was changed. - Fixed bug with wrong M-mode scaling for SmarUs and MicrUs device series. - Changed frequency settings of probe HL9.0/40/128-4 for LS128 devices. - Added support of SmartUs rev.C device with position sensor. - Updated settings of probe P4-2L20SI-6 for SmartUs rev.C,D devices. - Added support of Artus device. - New signature approved to drivers. - Added support of probe MC9-4R15S-3 for MicrUs rev.C devices. - Added support of probe L12-5L25NI-3 for SmartUs rev.B,C,D devcies. - Fixed synchronization bug for SmartUs rev.D devices. - Updated settings of probe MC10-5R10S-3 for MicrUs devices. - Added support of probe MCV9-5R10NI-3 for SmartUs rev.B devices. - Fixed bug with external synchronization for SmartUs rev.D devices. - Added compound mode for HL9.0/60/128-2 for LS128 devices. - Added bioeffect indices table for probe ID 0092 for SmartUs rev.C,D devices. Ver 1.16.7 2016 August 29 - Fixed bug with wrong frame size declaration for MicrUs devices. - Changed settings of probes L18-10L30H-4, C7-3R50NI-5 for SmartUs rev.C,D devices. - Changed geometry setting of probe MC10-5R10S-3 for MicrUs devices. - Updated settings of probes BIPL10-4L60N-4 and BIPC8-4R10N-4 for SmartUs rev.C,D devices. - Added support of probe ID 0094 for SmartUs rev.C,D devices. - Updated settings of probes C5-2R60N-3, MC10-5R10N-3, P4-2L20SI-3 for SmartUs rev.C,D devices. - Added support of pprobe P9-4L13SI-6 for SmartUs rev.C,D devices. - Updated bioeffect indices tables of probes BIPC9-4R10H and BIPL12-5L70H for SmartUs rev.C,D devices. - Changed settings of probes BIPL12-5L70H, BIPC9-4R10H for SmartUs rev.C,D devices. - Changed geometry of probe L7-3L40S-3 for Micrus rev.C,D devices. - Fxed bug with focal zones count restriction after probe change for SamrtUs rev.C,D devices. - Added support of probe ID 0092 for SmartUs rev.C,D devices. - Added support of probe L7-3L40S-3 for Micrus rev.C,D devices. - Changed settings of probes MCV9-5R10NI-3, MC10-5R20NI-3 for ClarUs devices. - Changed settings of probes MCV9-5R10NI-3, BIPC9-4R10H for SmartUs rev.C,D devices. - Added support of probe L15-6L25N-2 for ClarUs devices. - Added ext.synch. for SmartUs rev.D devices. - Added support of MicrUs rev.D devices. - updated settings and added bioeffect indices table of probe ID 0086 for SmartUs rev.C devices. - updated settings of probe P8-3L10SI-6 for SmartUs rev.C devices. - Updated settings of probe HL10.0/25/64D-3 for LS64 devices. - Updated settings of probe L18-11L13N-4 for SmartUs rev.C devices. - Fixed bug in ColorM scanning mode for SmartUs rev.D devices. - Added support of probe HL10.0/25/64D-3 for LS64 devices. - Added support of probe L18-11L13N-4 for SmartUs rev.C devices. - Updated settings of probe C7-3R50NI-5 for ClarUs devices. - Added support of probe T14-6L19S for Micrus devices. - Changed settings of probe P8-3L10SI-6 for SmartUs devices. - changed setting of probe P5-1L15SI-6 for SmartUs devices. - changed settings of probe P8-3L10SI-6 for SmartUs devices. - Added support of probe 3DC6-2R40NI-3 for SmartUs devices. - Updated settings of probe ID 0086 for SmartUs rev.C devices. Ver 1.16.6 2016 May 05 - Added support of probes ID 0086, MC5-2R20N-3, P4-2L20SI-3 for SmartUs rev.C devices. - Added support of probe C7-3R50S-3 for MicrUs rev.C devices. - Added support of probes L14-7L40N-2, C7-3R50NI-5 for ClarUs devices. - Added support of probe C3.5/60/128ZI-5 for Echo Blaster 128 devices. - Added support of probe C3.5/60/128ZI-5 for LS128 devices. - Added external freeze signal output for SmartUs rev.D devices. - Changed setting of probes L12-5L40N-3, P8-3L10SI-6, BIPC9-4R10H, MC10-5R6L for SmartUs rev.C devices. - Added bioeffect indices table for probes P8-3L10SI-6, BIPL10-4L60N-4, BIPC8-4R10N-4, BIPL12-5L70H, BIPC9-4R10H for SmartUs rev.C devices. - Changed bioeffect indices table for probes BIPL10-4L60N-4, BIPC8-4R10N-4 for SmartUs rev.C devices. - Added support of SmartUs Tomowave rev.D devices. - Chnaged settings of probe C7-3R50NI-5 for Clarus devices. - Changed settings of probe C3.5/60/128-3 for LS128 devices. - Fixed bug with packet size programming for SmartUs devices. - Changed geometry settings of probe PV6.5/10/128-3 for Echo Blaster 128 devices. Ver 1.16.5 2015 November 23 - Added support of probe MCV9-5R10NI-3 for SmartUs rev.C devices. - Updated settings of probe MCV9-5R10NI-3 for ClarUs devices. - Fixed minor bugs in SmartUs device drivers. - Implemented virtual convex scanning type for dual probes of Micrus Duo devices. - Updated settings of probe L15-6L25N-4 for SmartUs rev.C devices. - Fixed bug in installation script for SmartUs rev.C devices. - Added support of probe LVD8-4L65S-3 for MicrusDuo devices. - Changed some data for LS64 devices. - Added support of probes BIPL10-4L60N-4 and BIPC8-4R10N-4 for SmartUs rev.C devices. - Added support of probe MCV9-5R10NI-3 for ClarUs devices. - Changed settings of probe L15-6L25N-4 for SmartUs devices. - Changed settings of probe C7-3R50NI-5 for SmartUs rev.C devices. - Added support of probe P4-2L20SI-6 for SmartUs rev.C devices. Ver 1.16.4 2015 September 25 - Changed programming procedure of CW scanning mode for SmartUs devices. - Fixed bug with focal zone gain for probes L10-5L25NI-3 and L12-5L25NI-3 for ClarUs devices. - Updated settings of probe L18-10L30H-4 for SmartUs rev.C devices. - Added support of probe C7-3R50NI-5 for SmartUs rev.B devices. - Changed frequency management when scanning mode is changed for SmartUs devices. - Added support of probes C7-3R50NI-5 and P8-3L10SI-6 for SmartUs rev.C devices. - Added bioeffect indices settings for probe C3.5/60/128-3 for Echo Blaster 128 devices. - Changed settings of probe P5-1L15SI-6 for SmartUs rev.C devices. - Added support of probe L10-5L25NI-3 for ClarUs devices. - Added support of probe LV7.5/65/64-3 for LS64 devices. - Changed settings of probe C3.5/60/128-3 for LS128 devices. - Changed settings of probes L180-10L30H-4 and P5-1L15SI-6 for SmartUs rev.C devices. - Added support of probe C3.5/60/128-3 for Echo Blaster 128 devices. Ver 1.16.3 2015 July 23 - Added support of Smartus rev.C devices. - Fixed bug with probe detection of multiport Smartus devices. - Changed settings of probe CLA6.0/10/148 for SmartUs devices. - Changed settings of probe L12-5L25NI-3 for ClarUs devices. - Probe L10-5L25N-3 is renamed into L12-5L25NI-3 for ClarUs devices. - Added support of probe L3.5/180/128-3 for LS128 and Echo Blaster 128 devices. - Removed support of probe L5.0/80/128 for LS128 and Echo Blaster 128 devices. - Added support of probe HL9.0/25/64-3 for LS64 devices. - Fixed bug in installation script for probe HL9.0/60/128-2 for LS128 devices. - Added support of probe CASc-LAP-CA for LS128 devices. - Changed settings of probe P5-1L15SI-6 for SmartUs devices. Ver 1.16.2 2015 June 05 - Changed settings of probe MCV9-5R10N for ClartUs devices. - Added support of probe C15-7R7R-8 for ClarUs devices. - Updated settings of probe C5-2R60S-3 for MicrUs rev.C devices. - Added support of MicrUs Duo devices. - Added support of probe F6-5E192CR for SmartUs devices. - Added bioeffect indices table for probe LV7.6/60/128Z-2 for LS128 devices. - Updated parameters of probe LV7.6/60/128Z-2 for LS128 devices. - Changed coding of probes HL9.0/60/128-2, CASc-NUP-T40 for LS128 and Echo Blaster 128 devices. - Changed settings of probe HL9.0/60/128-2 for LS128 devices. Ver 1.16.1 2015 April 20 - Added support of probe C6.5/15/128-3 for LS128 devices. - Added support of probe MC4-2R20S-3 for MicrUs devices. - Fixed bug with firing pulses programing in Color Doppler modes for SmartUs devices. - Added support of probes L12-3L35N-7 and MCV10-4R9N-7 for SmartUs devices. - Added Doppler frequencies for probe L18-10L30H-4 for SmartUs devices. - Added doppler frequencies for probe L15-6L25N-4 for SmartUs devices. - Added support of probe HL9.0/60/128-2 for LS128 and Echo Blaster 128 devices. - Added support of probe L12-5L60N-2 for ClarUs and SmartUs devices. - Updated setting of probe L18-10L30H-4 for SmartUs devices. - Added bioeffect indices data of probe L18-10L30H-4 for SmartUs devices. - Changed power settings of probe L15-7L40H-5 for SmartUs devices. - Added support of steering and compound modes of scanning for probe BIPL7.0/60/128 for LS128 devices. - Added support of probe L15-6L25N-4 for SmartUs devices. - Added support of probe L15-6L25N-4 for ClarUs devices. - Changed geometry setting of probes C5-2R60S-3 and L12-5L40S-3 for MicrUs and MicrUs rev.C devices. - Changed geometry setting of probes HL9.0/40/64-3 and C3.5/60/64-3 for LS64 devices. - Fixed bug with ITHI frequency mode in M scanning. Ver 1.16.0 2015 February 23 - Updated UsgMon utility. - Added support of probes LV8-4L65S-3 and L15-6L25S-3 for MicrUs rev.C devices. - Added support of probe MC10-5R10N-3 for ClarUs and SmartUs devices. - Changed geometry settings of probe PV6.5/10/123 for LS128 devices. - Updated settings of probe L18-10L30H-4 for SmartUs devices. - Probe L18-10L30H is renamed into L18-10L30H-4. - Added support of MicrUs rev.C devices. - Added support of probe C5-2R60S-3M for MicrUs devices. - Added probe detection LOG output for SmartUs device driver. - Updated settings of probes BIPL12-5L70H and BIPC9-4R10H for SmartUs devices. - Probe P5-1L15SI is renamed into P5-1L15SI-6. - Changed frequency programming procedure to avoid BSOD for MicrUs devices. - Added bioeffect indices table for probe C5-2R60NI-5 for ClarUs evices. - Changed frequency settings of probe C5-2R60H-5 for SmartUs devices. - Updated settings of probes MC10-5R10N, C5-2R60N, L12-5L40N, P5-1L15SI, L15-7L40H-5 for SmartUs devices. - Probe MC10-5R20N-3 is renamed into MC10-5R20NI-3 (ClarUs and SmartUs devices). - Added support of test probe for MicrUs devices. - Changed frequency programming procedure for LS128, LS64 and ClarUs devices. - Changed log information output for SmartUs devices. - Added support of probe LV8-5L60N-2 for ClarUs devices. - Added support of probe MC10-5R20N-3 for SmartUs devices. - Corrected parameters of probe MC10-5R20N-3 for ClarUs devices. - Changed power supply monitoring algorihm for EB128 devices. - Updated settings of probe MC10-4R12N-4 for SmartUs devices. - Updated filter settings of probes L12-5L40S-3, L12-5L32S-3, C5-2R60S-3, MC10-5R10S-3, MCV9-5R10S-3, MC8-4R20S-3 and T10-5L17S for MicrUs devices. - Updated settings of probe PV6.5/10/1258-3 for LS128 devices. - Updated settings of probe LV8-5L60N-2 for SmartUs devices. - Added support of probe LV8-5L60N-2 for SmartUs devices. - Added support of probe MCV9-5R10N-3 for SmartUs devices. - Added support of probe MC10-5R10N for SmartUs devices. - Updates settings of probe MC10-5R6L for SmartUs devices. - Added support of probe C5-2R60NI-5 for ClarUs devices. - Added support of probe C5-2R60NI-5 for SmartUs devices. - Implemented new gain compensation algorithm for SmartUs devices. - Changed programmming of TGC property of SmartUs devices (not stops scanning during programming procedure). - Updated setting of probes L12-5L60N and MCV9-5R10N for SmartUs devics. - Added possibility to output scan state signal for Clarus devices. - changed settings of probe PV6.5/10/128-3 for LS128 devices. - changed settings of probe MC8-4R20S-3 for MicrUs devices. - changed transmit power managment for MicrUs devices. - Added support of probes L12-5L32S-3, C5-2R50S and T10-5L17S for MicrUs devices. Ver 1.15.2 2014 November 19 - Updated settings of probes BIPC9-4R10H and BIPL12-5L70H for SmartUs devices. - Added support of probes L15-10L25S-3 and MC8-4R20S-3 for MicrUs devices. - Implemented ITHI frequency mode for SmartUs devices. - Updated settings of probes C5-2R60HI-5 and L18-10L30H for SmartUs devices. - Added bioeffect indices table for probe C5-2R60HI-5 for SmartUs devices. - Fixed bug in transmit power programming. - Changed algorithm of transmit power programing for MicrUs devices. - Changed settings of probe C5-2R60S-3 for SmartUs devices. - Added possibility to change gain range for MicrUs devices. - Changed settings of probe L12-5L40S-3 for MicrUs devices. - Updated focues setting of probe C5-2R60HI-5 for SmartUs devices. - MicrUs device drivers are updated to support medium and standard line densities. Ver 1.15.1 2014 October 16 - Blocked possibility to select phantom volume in less than 3 mm depth for HPRF in PW mode on SmartUs devices. - Fixed bug with audio output in B+PW mode on SmartUs devices. - Fixed bug with HPRF PW mode programming for SmartUs devices. - Changed procedure of frequency adjust for SmartUs device driver (BSOD during frequency change). - Added support of probe PV6.5/10/128-3 for LS128 beamformer. - Added support of probe PV6.5/10/128-3 for Echo Blaster 128 beamformer. - Added support of SmartUs-3Z beamformer. - Added support of MicrUs scanners. - Improved management of control request processing in SmartUs device driver. - Changed settings of probes L12-5L40N, L12-5L40N-4, L12-5L40H-5, C5-2R60HI-5, C5-2R60H-5, P5-1L15SI for SmartUs devices. - Added bioeffect indices tables of probes C3.5/60/64-3. HL9.0/40/64-3 for LS64 devices. - Changed settings of probe MC10-5R6L for SmartUs beamformer. - Changed programming of CW gain for SmartUs beamformer. - Smartus device driver updated for firmware ver. 3.xx support (revision 3.01) - Fixed bug with wrong declaration of ultrasound data size for SmartUs devices. Bug can cause ultrasound data processing and rendering stop. - Added support of probe MC10-5R6L for SmartUs devices. - Updated settings of probes C5-2R50N-4 and C5-2R60HI-5 for SmartUs devices. - Fixed bug with THI scanning in Color Doppler modes for SmartUs devices. - Added bioeffect indices for probe C5-2R60H-5 for SmartUs devices. - Updated settings of C5-2R60H-5 probe for SmartUs devices. - Added support of Power Doppler modes for SmartUs devices. - Added bioeffect indices for probe P5-1L15SI for SmartUs devices. - Updated settings of P5-1L15SI and L12-5L40N-4 probes for SmartUs devices. - Updated settings of C5-2R50N-4 probe for ClarUs devices. - Changed acoustic output power settings of probe EC6.5/10/128 for LS128 device in triplex mode. - Added connection error loging for SmartUs device drivers. - Added support of PW Spectral averaging processing for SmartUs devices. - Fixed bug with incorrect TIC report when no indices are defined for SmartUs devices. Ver 1.15.0 2014 May 27 - Updated settings of probe P5-1L15SI for SmartUs device. - Fixed initilization of focus settings bug in SmartUs device driver. - Added support of test probe for SmartUs devices. - updated settings of probe L15-7L40H-5 for SmartUs devices. - Lower PRF value of PW/CW modes for SmartUs devices is changed to 1 KHz. - updated settings of probes C5-2R60N, CLA6.0/10/148, L12-5L40N-4 for SmartUs devices. - Added bioeffect indices table support for probes L15-7L40H-5, CLA6.0/10/148, L12-5L40N-4 for SmartUs devices. - Fixed bug with incorrect programming of convex probes parameters for SmartUs devices. - Updated settings of probe CLA6.0/10/148 for SmartUs device. - Added support of probe C5-2R60H-5 for SmartUs devices. - disable bioeffect indices for some probes for LS128 and LS64 devices. - Fixed bug with correct PRF change in triplex mode for SmartUs devices. - Added support of probes C3.5/20/128-3, C3.5/60/128-3, EC6.5/10/128-3, HL9.0/40/128-3 for LS128 devices. - Added support of MC10-4R12N-4 probe for SamrtUs devices. - Updated settings of probe L7-3L80N for SmartUs devices. - Added support of Echo Blaster 128 with 6-bits probe connector. - Added support of C3.5/20/128-3 probe for Echo Blaster 128 devices. - Fixed bug with Color Doppler timers programming in triplex modes. - Added support of LS128-3Z devices. - Added support of C3.5/60/80-7 probe for LS128 devices. Ver 1.14.3 2014 March 05 - Updated settings of MC10-5R10N probe for ClarUs devices. - Accurate definition of hardware registers in source of SmartUs driver (not for release readme). - Updated settings of MC10-4R12N-4, MC10-5R20N-3, L10-5L25N-3 probes for ClarUs devices. - Updates settings of probe L15-7L40H-5 for SmartUs devices. - Added support of probe L10-5L25N-3 for ClarUs devices. - For SmartUs devices all linear probes with 192 elements names suffix 'HI' changed into 'H'. - Added support of probes C3.5/20/64-3, EC6.5/10/64-3, PV6.5/10/64-3 for LS64 devices. - Updated settings of probe HL9.0/40/128-4 for Echo Blaster 128 devices. - Updated settings of probe HL9.0/40/128-4 for LS128 devices. - Updated settings of probe L12-5L40N-4 for ClarUs devices. - Added support of probes MC10-4R12N-4, MC10-5R20N-3 for ClarUs devices. - Changed power reference for probe BIPC6.5/9/128Z for LS128 devices. - Added support of probe MCV9-5R10N for SmartUs devices. - Added possibility of limit output power range for Echo Blaster 128 devices. - Added some check to validate incoming USB data for SmartUs devices. - Updated settings of probe L7-3L80N for SmartUs devices. - Added support of probe MC5-2R20N-3 for ClarUs devices. - Added support of probe MCV9-5R10N-3 for ClarUs devices. - Changed driver for SmartUs devices to support FPGA EEPROM write. - Removed noise during scanning parameters switch in PW mode for SmartUs devices. - Added support of probe L12-5L40N-3 for SmartUs and ClarUs devices. - Added support of probe HL9.0/40/64-3 for LS64 devices. Ver 1.14.2 2013 December 31 - Addes support of probe C5-2R60HI-4 for SmartUs devices. - Changed settings og probe L15-7L40HI-5 for SmartUs devices. - Fixed bug with focus programming for SmartUs devices. - Added support of probe C3.5/60/128-6 for LS128 devices. - Added support of probe C5-2R50N-4 for ClarUs and SmartUs devices. - Added support of probe C5-2R60N-3 for ClarUs and SmartUs devices. - Added support of probe HL9.0/40/128-4 for Echo Blaster 128 and LS128 devices. - Changed probe naming algorithm. - Added support of C3.5/60/64/SR probe for LS64 devices. - Corrected delays of BIPL7.0/60/128 probe for LS 128 device. - changed apodization filter for L18-10L30HI, L12-5L40HI, L15-7L25N, L15-7L40HI, L12-5L40HI/VM probes for SmartUs device. - Changed settings of P5-1L15SI probe for SmartUs devices. - Added support of L18-10L30HI probe for SmartUs devices. - Added support of CW mode for SmartUs devices. - New Altera firmware for SmartUs devices. Changes in AFE, TFC and Bandpass filters. - Added support of L12-5L40N/VM probe for ClarUs devices. - Added support of L12-5L40N/VM and L12-5L40HI/VM probes for SmartUs devices. Ver 1.14.1 2013 October 31 - Updated power settings and bioeffect indices of C7-3R50N/BS probe for ClarUs devices. - Updated power settings and bioeffect indices of C4.5/50/128/BS probe for LS128 devices. Ver 1.14.0 2013 October 18 - Added pulser voltage log output for following devices: LS64, LS128, ClarUs, Echo Blaster 128. - Added support of probe BIPC6.5/9/128Z for LS128 device. - Removed support of probe UT-LA5.0/128 for Echo Blaster 128 devices. - Added support of probe LV7.5/60/128Z/BS for Echo Blaster 128 devices. - Updated probe LV7.5/60/128Z/BS settings for LS128 devices. - Fixed bug with PW power reference calculation in SmartUs device driver. - Added support of probe LV7.5/60/128Z/BS for LS128 devices. - Removed support of CLA3.0/20/96 probe for Echo Blaster 128 devices. - Removed support of CLA3.0/20/96 probe for LS128 devices. - Added support of C4.5/50/128/BS probe for Echo Blaster 128 devices. - Added support of C4.5/50/128/BS probe for LS128 devices. - Updated settings for SmartUs device probes. - Added support of probe CLA6.0/10/148 for SmartUs device. - Updated settings of probes L12-5L40HI, C5-2R60HI, C7-5R50HI, L12-5L40N, C7-3R50N, C5-2R60N/BS, P5-1L15SI, L15-7L40HI for SmartUs devices. - Added support of probes C7-3R50N/BS, C5-2R60N/BS, L12-5L40N/BS for SmartUs device. - probes C7-3R50N/B, C5-2R60N/B, L12-5L40N/B are renamed into C7-3R50N/BS, C5-2R60N/BS, L12-5L40N/BS. - Added support for LS128 with 6 bits probe connectors. - Fixed bug with pulser voltage for triplex mode in SmartUs devices. - Added frequency 12 MHz for Clarus probes SP12-5L/B/S. - Added installation script for SmartUs beamformer. - Added support of C7-3R50N/B, C5-2R60N/B and L12-5L40N/B probe for ClarUs devices. - Added support of SP12-5 probe for ClarUs devices. - Implemented special steering for SP12-5 probe. - Implemented compound slop scanning mode for LS128 and ClarUs devices. - Inproved sound quality of PW mode for LS128, LS64 and ClarUs devices. Ver 1.13.7 2013 April 09 - Updated setting of BIPL7.0/60/128 probe for LS128 devices (PW power reference, ColorDoppler transmit delays, bioeffect indices table). - Fixed bug in frequencies setting of ClarUs device for probes L12-5L60N, L4-1L180L, L15-7L25N, L15-6L25N. - Added support of loging of AcuVista device to ultrasound scanner monitor. - Added possibility to output freeze synchronize signal to Echo Blaster 128 device. Ver 1.13.6 2013 February 11 - Added synchronization of sacnning properties programing. - increased USB transfer buffer size for LS64 devices to improve transfer stability. - Fixed bug with hardware programming that can cause scanning hold in Triplex modes for LS128, LS64 and ClarUs devices. - Fixed bug in optimal output acoustic power calculation for LS128 and ClarUs devices. - Added SmartUs logs support for monitor utility. - Updated settings of HL12.0/25/128 probe for LS128 devices. - Fixed bug with wrong output frame size report for LS128, LS64, ClarUs devices. Ver 1.13.5 2012 October 18 - Fixed bug in installation script with correct LS64-2T installation. - Updated power settings for L15-6L25N, L15-7L25N probes for ClarUs devices. - Added bioeffect indices for L15-6L25N, L15-7L25N probes for ClarUs devices. Ver 1.13.4 2012 September 28 - Fixed bug with bioeffect indices calculation for PW mode for LS64 devices. - Fixed bug with bioeffect indices calculation for PW mode for LS128 and ClarUs devices. - changed power setting values of probe HL9.0/40/128 for LS128 devices. - adjusted correct values of bioeffect indices and power reference for SP12-5 probes for LS128 devices. - Adder support for LS128-3Z devices. - Added support of SP12-5 probe for LS128 devices (1st implementation). - Added support of steering and compound scanning for probe BIPL7.5/70/128 for LS128 devices. - Updated bioeffect indices table of L12-5L40N probe for ClarUs and ClarUs4D devices. - Added support of compound modes for LS128 and ClarUs devices. Ver 1.13.3 2012 April 03 - Changed frequency settings of L3.5/180/96 probe for LS128 devices. - Changed frequency settings of L4-1L180L probe for ClarUs devices. Ver 1.13.2 2012 March 29 - Fixed bug with preset of acoustic output power in B-modes for LS128 and ClarUs devices. Ver 1.13.1 2012 March 23 - Added support of probe UT-C5.0R20/128 for LS128 device. - Added support of probe L15-6L25N for ClarUs device. - Added possibility to preset acoustic output power in B-modes for LS128 and ClarUs devices. Ver 1.13.0 2012 March 19 - Updated bioeffect indices table of BIPL7.0/60/128 probe for LS128 devices. - Updated bioeffect indices table of C3.5/20/128 probe for LS128 devices. - Updated output acoustic power settings of BIPL7.0/60/128 and C3.5/20/128 probes for LS128 devices. - Updated output acoustic power settings of probe L12-5L40N for ClarUs and ClarUs4D devices. - Updated output acoustic power settings of probe L15-7L25N for ClarUs devices. - Disabled view area change for L12-5L40N probe for ClarUs and ClarUs4D devices. - Disabled view area change for L15-7L25N probe for ClarUs devices. - Updated bioeffect indices table of C7-3R50N probe for ClarUs and ClarUs4D devices. - Updated bioeffect indices table of C4.5/50/128 probe for LS128 devices. - Updated bioeffect indices table of L12-5L60N probe for ClarUs and ClarUs4D devices. - Updated output acoustic power settings for L12-5L60N probe for ClarUs and ClarUs4D devices in Triplex and PW modes. - Updated indices table of CASc-NUP-T40 probe for LS128 device. - Enhanced dynamic range available values for LS64, LS128 and ClarUs devices. - Fixed bug with probe connection identifier for ClarUs4D devices. - Implementer support of probe EEPROM property for probe objects for ClarUs4D devices. - Updated power reference of L15-7L25N probe for ClarUs devices. - Updated frequency settings of BIPL7.0/60/128 probe for LS128 device. - Fixed installation bug for BIPL7.0/60/128 probe LS128 device driver). - Added support of BIPL7.0/60/128 and CASc-NUP-T40 probes for LS128 devices. - Updated settings for L15-7L25N probe for ClarUs devices. - Added support of probe L15-7L25N for ClarUs device. - Added support of steering and compound scanning types for L12-5L40N and L15-7L25N probes for ClasUs devices. - Added support of steering and compound scanning types for HL9.0/40/128 probe for LS128 and LS64 devices. - Implemented Steering B scanning and Compound Scanning for LS128 and ClarUs devices. Ver 1.12.13 2011 September 20 - Changed power referencies of 3D6-3R40N probe for ClarUs4D devices. - Updated bioeffect indices for HL9.0/60/128 probe for LS128 devices. - Updated output acoustic power settings for HL9.0/60/128 probe for LS128 device in Triplex mode. - Changed data transfer protocol for ClarUs4D devices. - Changed GearRatio parameter for 3D probe (was 40, now 45). - Changed maximum 3D view angle for 3D probe (was 85, now 75). - Changed supply voltage check algorithm to initialize Echo Blaster 128 beamformer device. Ver 1.12.12 2011 June 27 - Changed management of volume frames in 4D mode for ClarUs4D devices. - Updated settings of 3D6-3R40N probe for ClarUs4D devices (linear motor board compatible). - changed proxy module to avoid dependencies from VC run-time libraries. - Fixed minor bugs in ClarUs4D device drivers. - Added support of ClarUs4D-2M device. - Added support of L4-1L180L probe for ClarUs device. - Added support of L3.5/180/96 probe for LS128 device. - Add support of ClarUs2M device. - Added support of ClarUs4D device. - Added support for LS64-2T devices. - Fixed bug in ClarUs4D device driver (BSOD when EchoWave2 enters in 4D mode) Ver 1.12.11 2011 May 23 - updated installation script to work correctly with old devices. - updated indices tables of biplanar probe for LS128 beamformer. - updated HL10.0/25/96 probe settings for LS128 beamformer: output acoustic power reference available Color Doppler lines density range disabled ViewArea bioeffect indices - Fixed bug with management of several devices connected to the system simultaneously for LS128 and ClarUs devices. - Pull time to check connected probe for ClarUs changed from 160ms to 1s. - Removed support of HC3.5/60/128 probe for Echo Blaster 128 device. - Added support of L3.5/180/96 probe for EchoBlaster 128 device. - Added support of L12-5L25L probe for ClarUs devices. - Probe 3D7-1R40N renamed to 3D6-3R40N. - Added possibility to preset number of available line densities for Color Doppler modes for LS128 and ClarUs devices. - Changed acoustic output power settings for biplanar probes (BIPC6.5/10/128 and BIPL7.5/70/128) for LS128 devices. - Added support of probe LT10-5L17N probe for ClarUs device. Ver 1.12.10 2011 February 15 - Corrected PW sample volume position calculation for ogicScan 128 and ClarUs devices. - Corrected settings of HL12.0/25/128 probe for LS128 device. Ver 1.12.9 2011 January 21 - Added battery status property support for ClarUs and LS64 device drivers. - fixed bug with bioeffect indices calculation for CalrUs, LS128 and LS64 device in B and M modes. - updated indices of BIPL7.5/70/128 and BIPC6.5/10/128 probes for LS128 devices. - changed manifest of drv_uninst utility. Ver 1.12.8 2010 November 23 - Fixed bug in HL9.0/60/128 probe settings for LS128 devices. - Changed supply voltage range to turn ClarUs devices on. - added support for L5.0/40/128 probe - Fixed bug in frequency settings for probes L4-1L170L and L2.5/170/96. - Removed support of probe L4-2L170L for ClarUs devices. - Added support of probe L4-1L170L for ClarUs devices. - Added support of C5.0/20/96 probe for LS128 devices. - Changed AGC curve of 6MHz multifocus setting for following LS128 probes: EC6.5/10/128, PV6.5/10/128, EC7.0/10/128 - Changed AGC curve of 6MHz multifocus setting for following LS64 probes: EC6.5/10/64, PV6.5/10/64 Ver 1.12.7 2010 August 13 - Fixed bug in installation script (logicscan.inf does not match for image in CAB file) Ver 1.12.6 2010 July 09 - Changed installation script for LS128 device drivers to allow automatic install of non-composite device revisions. - updated power reference values for LS64 probes. Ver 1.12.5 2010 June 04 - Fixed bug with optimal frame rate calculation for Color Doppler Modes for LS128, LS64, ClarUs devices. - Updated Power reference values for some probes (C3.5/60/128, C3.5/20/128, C4.5/50/128, PV6.5/10/128, C2.5/60/128, LV7.5/60/96) - Fixed bug with supply voltage management when device is resume after sleep for LS128, LS64, ClarUs devices. Ver 1.12.4 2010 May 28 - Added support of T9.0/19/128 probe for LS128 devices. - Fixed bug with system restore after power supply goes normal for LS128, LS64, ClarUs, Echo Blaster 128 devices. - Changed supply voltage monitoring range for ClarUs devices (low voltage 10-12 Volts) - added support of L13-5L25N probe for ClarUs device Ver 1.12.3 2010 May 12 - updated range of valid supply voltages for ClarUs devices (10-15 volts). Ver 1.12.2 2010 May 07 - updated installation package (install with no console windows and uninstall system drivers files) - Fixed bug in LS128 and Clarus drivers (some low Color Doppler lines density and low PRF combination causes BSD). - Fixed bug with probe BIPC6.5/10/128 for LS128 device settings. - added support of L2.5/170/96 probe for LS128 device - changed installation script for Echo Blaster 128 devices - fixed bug with acoustic power-level overshoot in PW modes for LS128 Rev.A devices. - removed support of LA3.5/2x48 probe for Echo Blaster 128 device - added support of L2.5/170/96 probe for Echo Blaster 128 device Ver 1.12.1 2010 February 09 - fixed bug in installation script (Windows7 automatic installation) - added output signal control to indicate scanning state for LS128 devices. - addes support of bioeffect indices for LS64 devices - probe frequency control implemented inside LS64 device driver Ver 1.12.0 2010 January 15 - added 64-bit support for all device drivers. - all executable, catalog and system files are digitally signed. - fixed bug with PRF exceeding in PW mode for LS64 devices. - added support of test probe for ClarUs device. Ver 1.11.1 2009 November 26 - updated frequencies for L12-5L40N, C7-3R50N and C5-2R60N probes for CarUs device. - Fixed bug with USB streaming transfer termination for Echo Blaster 128 and Echo Blaster 64 devices. - Fixed acoustic power overriding for HPRF PW mode for LogisScan 128 and ClarUs devices. - HPRF mode implemented for LS128 Rev.B and ClarUs devices. - Implemented new probe detection method for ClarUs devices. - Fixed bug in beamformer drivers related with driver object reopen. Ver 1.11.0 2009 September 18 - updated settings of EC6.5/10/128 probe for LS128 - added installation of ClarUs device drivers. - Added monitoring of supply voltage for Echo Blaster 128 devices. - Added monitoring of pulser voltage for LS128 devices. - updated monitor utility to collect log information from ClarUs device - added support of probe L12.0/40/128 for LS128 - added support of triplex modes for LS64 Rev.B devices Ver 1.10.8 2009 June 29 - removed support of N3.0/140/32 probe for Echo Blaster 64. - added support of HL9.0/40/64 probe for Echo Blaster 64. - fixed bug in system suspend and resume handler for LS128 and LS64 devices - added support of bioeffect indices for EchoBlaster 128 device driver - Changed list of available values for Doppler PacketSize for LS128 Rev.B (8,10,12,16). Ver 1.10.7 2009 May 19 - added support for LA3.5/2x48 probe for Echo Blaster 128 devices Ver 1.10.6, 2009 April 23 - Fixed power supply checking bug during system resume in LS128 device driver. - Changed list of available values for Doppler PacketSize for LS128 Rev.B (8,10,16). - Implemented enhanced PRF for B+PW mode for LS128 Rev.B devices. - updated frequency settings for BIPC6.5/10/128 and BIPL7.5/70/128 probes for LS128 devices. - added support for probe BIPL7.5/70/128 for LS128 devices - changed settings for probe BIPC6.5/10/128 for LS128 devices Ver 1.10.5, 2009 March 23 - updated frequency settings for C2.5/60/128 for LS128B - updated frequency settings for C5.0/15/128 for LS128B Ver 1.10.4, 2009 March 12 - Fixed bug in LS64 installation script (old SYS file was not replaces with new file automatically). - PW signal is blanked below 1,2% of signal level for LS64 and LS128 device families. - changed frequency settings for following probes for LS128 Rev.B devices: C2.5/50/128, C5.5/50/128 - add support for LS64 Rev.B - added support for EC7.0/10/128 probe for EchoBlaster128 device. - added support for EC7.0/10/128 probe for LS128 device. - updated focus data files for EC8.0/10/128 probe for LS128 device. - changed power reference values for C5.0/15/128 probe for LS128 device. - changed power reference values for C5.5/50/128 probe for LS128 device. Ver 1.10.2, 2009 January 29 - removed blank ultrasound points in PW mode below 2.5% of signal level for LogisScan 64 and LS128 device families. Ver 1.10.1, 2008 December 29 - updated firmware RBF for LS128 Rev.B device. Improved Color Doppler data quality. - updated 'Ultrasound Scanner Monitor' utility to handle logs of firmware update utilites. - fixed bug in LogisScan 64 device with delays changing after lines density change - updated Echo Blaster installation scripts to force Kernel Stream modules installing when device drivers are installed - Increased power for some probes for LS128 Rev.B in ColorDoppler modes. - Added support for Bioeffect Indices (MI, TIC, TIS, TIB) for LS128 device. - added support of HL12.0/25/128 and C5.5/50/128 probes for LS128 Rev.B Ver 1.10.0, 2008 November 10 - added support for THI modes for LS128 Rev.B devices - updated Doppler Power reference for C2.5/60/128 probe for LS128 Rev.B device. - fixed bug with scanning hold in triplex modes for LS128 Rev.B device. - changed range of valid PDI signal scale values for LS128 Rev.B devices - added settings for C5.0/15/128 probe for LS128 device - updated firmware RBF for LS64 device (improved HP filter for PD) Ver 1.9.1, 2008 September 18 - fixed bug with Color Doppler calculation time registers overflow in LS128 Ver 1.9.0, 2008 September 10 - added support for LS128 Rev.B - changed automatic focus select algorithm for Doppler modes for LS64 and LS128 devices Ver 1.8.1, 2008 June 25 - updated settings for probe L3.5/170/96 for LS128 and Echo Blaster 128 devices Ver 1.8.0, 2008 May 16 - removed power limit for Color Doppler modes for LS128 and LS64 devices. - updated firmware RBF for LS128 and LS64 devices to currect sound in PW modes. - implemented system power management support for LS64 and LS128 devices. - added support of L5.0/80/128 probe for LS128 device. Ver 1.7.0, 2008 March 19 - updated firmware RBF file to increase quality of PW data for LS64 and LS128 devices - implemented Spectral Signal Averaging feature for LS64 and LS128 devices - fixed bug with wrong generation of B and CD formats for LS128, LS64, EchoBlaster 64 and EchoBlaster 128 devices Ver 1.6.2, 2008 March 13 - fixed minor bug (USB bus initialization) in LS64 and LS128 drivers. - added support for composite LS128 devices . - added firmware registers reading feature for LS128 devices. - added support of HL9.0/60/128 probe for LS128 device. Ver 1.6.1, 2008 February 28 - changed power monitor settings for LS128 device - changed log output sequence in USMON.EXE utility - fixed bug in proxy module (memory access violation) Ver 1.6.0, 2008 February 12 - implemented ColorDoppler pulse test mode Ver 1.5.7, 2008 February 08 - added support of HL9.0/40/64 probe for LS64 device - updated minimum Color Doppler value for LS128 and LS64 devices - fixed bug in LS64 firmware RBF file - updated firmware RBF file for LS128 device - added support of HL10.0/25/96 probe for LS128 device Ver 1.5.6, 2008 January 31 - implemented power monitor for LS128 and LS64 devices - enhanced Doppler Wall Filter for LS128 and LS64 devices Ver 1.5.5, 2008 January 23 - added support of L3.5/170/96 probe for LS128 device - updated PW power reference values for LS128 and LS64 devices - updated range of available values for PW Scale, PW Audio, PDI/DPDI Scale properties for LS128 and LS64 devices Ver 1.5.4, 2008 January 18 - updated RBF files for LS64 and LS128 devices (wall filter improved) Ver 1.5.3, 2007 December 28 - Fixed bug with PRF range in B+PW mode for LS128 and LS64 devices Ver 1.5.2, 2007 December 22 - Fixed PW sample volume position bug - Fixed bug with B+PW mode timing for LS128 and LS64 devices - Updated RBF files for LS128 and LS64 device - Updated range of available Doppler frequencies for LS128 and LS64 devices - Added support for EC8.0/10/128 probe for LS128 device. Ver 1.5.1, 2007 December 17 - Added support for C2.5/60/128 probe for LS128 device. Ver 1.5.0, 2007 December 05 - added Doppler multifrequency support for LS128 & LS64 devices - added Doppler PRF change notifications for LS128 & LS64 devices Ver 1.4.1, 2007 November 21 - improved Color Doppler (CFM/PDI/DPDI) modes scanning rate for LS64. Ver 1.4.0, 2007 November 10 - improved Color Doppler (CFM/PDI/DPDI) modes scanning rate for LS128. - added B+PW and PW modes support for LS128 & LS64 devices Ver 1.3.1, 2007 October 09 - added support of L3.5/170/96 probe for LS128 device Ver 1.3.0, 2007 September 27 - added support of C5.0/15/128 probe for EchoBlaster 128 device Ver 1.2.4, 2007 September 21 - added support for LS128-2Z devices - fixed bug with second connector initialization for Echo Blaster 128-2Z devices - fixed minor bugs in proxy module Ver 1.2.3, 2007 September 17 - fixed bug in LS64 device driver with ViewArea on low lines density - added support of C5.0/15/128 probe for LS128 device Ver 1.2.2, 2007 July 31 - updated UsgMon utility - updated M-mode scan-converter Ver 1.2.1, 2007 July 16 - Corrected system date -related bug inside "System Information" and "Ultrasound Scanner Monitor" tools. - Updated renderer filter to support static bitmap output. Ver 1.2.0, 2007 July 10 - implemented Spatial Filter for Color Doppler modes - implemented Frame Averaging Filter for Color Doppler mode - drivers for LS64 and LS128 devices updated to support 5.33 MHz frequency for Color Doppler modes - updated hardware initialization file (RBF) for LS128 device driver - updated M-mode scan-converter Ver 1.1.0, 2007 July 05 - added Teleview utility - fixed bug with time stamp for M-mode output samples - updated installation files to install system multimedia drivers together with device drivers Ver 1.0.10, 2007 June 18 - updated delay files for all LS64 device probes Ver 1.0.9, 2007 May 29 - updated delay files for HL7.5/40/64 probe (LS64 device) Ver 1.0.8, 2007 May 17 - added device driver for LS64 beamformer Ver 1.0.7, 2007 May 16 - updated USGFW libraries Ver 1.0.6, 2007 May 14 - added support for OEM probe N3.0/140/32 for EchoBlaster 64 device Ver 1.0.5, 2007 April 24 - updated band-pass gidital filter for LS128 device - updated monitor application Ver 1.0.4, 2007 April 16 - improved acoustic power limitation for LS128 beamformer - added log to indicate wrong USB bus configuration Ver 1.0.3, 2007 April 13 - implemented acoustic power limitation for LS128 beamformer - improved stability of LS128 scanning processes Ver 1.0.2, 2007 April 05 - fixed bug with special hardware control (Lithotripsy) - added special signal output for Echo Blaster 128 - added dynamic focus support for T-probes for Echo Blaster 128 - fixed bugs with convolution filtering in scan-converter - T-probes are renamed in EB128 driver service registry key - fixed bug in EB64 driver with correct geometry calculation for linear probes - fixed bug in EB128 driver with capabilities report for T-probes Ver 1.0.1, 2007 March 16 - fixed bug with surprise removal in LS128 driver - added events logging (UsMon.exe application) for all drivers - fixed bug with test mode switch in Echo Blaster 64/128 drivers Ver 1.0.0, 2007 February 28 - initial release 4. TROUBLESHOOTING See corresponding chapter of device User Guide. (c)1992-2024 TELEMED Internet: www.pcultrasound.com, www.telemed.lt Information e-mail: info@pcultrasound.com, info@telemed.lt TechSupport e-mail: support@pcultrasound.com, support@telemed.lt